strain gage animation

Japanese

Dynamic Strain Amplifiers

Dynamic Strain Amplifier

Features

    AC Strain Amplifiers Models 5683 / 5684

  • Excellent in high sensitivity, high stability and in noise-resistant capability ensuring for measurement in the use of strain gages.
  • Model 5683 uses the frequency of 5kHz in its bridge power supply, which ensures the response characteristics from DC to 2kHz.
  • Model 5684 uses the frequency of 2kHz in its bridge power supply, which ensures the response characteristics from DC to 10kHz.
  • An automatic volume elimination function is installed on the balance circuit.
  • A bridge circuit disconnection checking function is equipped on the system, which enables an easy trouble-free operation during measurement (common to Model 5693).
  •  

    DC Strain Amplifier Model 5693

  • Suitable for a high accuracy measurement with the use of strain gage type transducers (for measurement of load, pressure, torque, acceleration and so on).
  • A stabilized voltage of 2, 3, 5, 9 or 10 V is used in its bridge power supply, which realzes a hircuit higher respons frequency of DC to 500kHz.
  • An auto balance function is mounted on the balance circuit, which enables to perform initial balancing operations instantaneously with high accuracy.

Specifications

Item AC Strain Amplifier Model 5683
(General Purpose Type)
AC Strain Amplifier Model 5684
(High Response Type)
DC Strain Amplifier Model 5693
(Wide Bandwidth Type)
Number of Channels 1 channel/unit
Bridge resistance 60Ω to 1kΩ
Gage Factor 2.00
Bridge power voltage supply AC0.5V,2V 5kHz sine wave. Sync input/output signal 2.5VAC AC0.5V,2V 25kHz sine wave. Sync input/output signal 2.5VAC DC 2V,3V,5V,9V,10V
Disconnection check function

Detecting disconnection and short of input bridge circuit and displaying checked result by LED

Balance adjusting range(Auto-balance) Deviation of resistive value:±2%(±10,000×10-6 in strain)
Capacitive imbalance:Approx.2,000pF
Deviation of resistive value:±2%(±10,000×10-6 in strain)
Balance adjusting accuracy Within ±0.4×10-6 in strain
(RANGE=200,Without FINE,BV=2V)
Within ±1×10-6 in strain
(RANGE=500,Without FINE,BV=2V)
Within ±1×10-6 in strain
(RANGE=1k,without FINE,BV=2V)
Maximum input range ±200,000×10-6 in strain
(RANGE=20k,FINE=×2.5,BV=0.5V)
±500,000×10-6 in strain
(RANGE=50k,FINE=×2.5,BV=0.5V)
±125,000×10-6 in strain
(RANGE=50k,FINE=×2.5,BV=2V)
Voltage sensitivity ±10V within ±200×10-6 in strain
(RANGE=200,without FINE,BV=2V)

±10V within ±500×10-6 in strain
(RANGE=500,without FINE,BV=2V)

±10V within ±1000×10-6 in strain
(RANGE=1k,without FINE,BV=2V)
Measurement range change 200,500,1k,2k,5k,10k,20k(×10-6 in strain,×2/BV in value),OFF 500,1k,2k,5k,10k,20k,50k(×10-6 in strain,×2/BV in value),OFF 1k,2k,5k,10k,20k,50k(×10-6 in strain,×2/BV in value),OFF
Fine adjustment Continuously chnangeable in FINE RANGE,2step changing amount ca be selected
Internal calibrator Set value: ±1 to ±9,999×10-6 in strain,
Accuracy: ±(0.5%rdg+0.5×10-6 in strain)
Set value: ±1 to 9,999×10-6 in strain
Accuracy: ±(0.2%rdg+0.5×10-6 in strain)
Nonlinearity ±0.1%FS ±0.2%FS ±0.01%FS
Frequency response DC to 2kHz±10% DC to 10kHz±10% DC to500kHz、+1,-3dB
High pass filter 0.5Hz: 2-pole Butterworth type
(Filter descent response: -12dB/oct)
0.5Hz: 2-pole Bessel type
(Filter descent response: -12dB/oct)
Low-pass filter 10,30,100,300,500Hz 4-pole Butterworth type
(Filter descent: -24dB/oct)
10,30,100,500Hz,3kHz 4-pole Butterworth type
(Filter descent: -24dB/oct)
10,30,100,1k,30kHz 4-pole Bessel type
(Filter descent: -24dB/oct)
Stability Zero drift: Within ±0.1×10-6 in strain/℃,Within ±0.5×10-6 in strain/24h
Sensitivity: Within ±0.05%/℃,within ±0.2%/24h
Zero drift: Within ±1×10-6 in strain/℃以内,within ±5×10-6 in strain/24h
Sensitivity: Within ±0.01%/℃,within ±0.05%/24h
Noise level 2.0×10-6 in strain p-p RTI(W/B,RANGE=200,Without FINE,BV=2V,120Ω in bridge)
0.6×10-6 in strain p-p RTI(DC to 100Hz,RANGE=200,Without FINE,BV=2V,120Ω in bridge)
6.0×10-6 in strain p-p RTI(W/B,RANGE=500,Without FINE,BV=2V,120Ω in bridge)
2.0×10-6 in strain p-p RTI(DC to 100Hz,RANGE=500,Without FINE,BV=2V,120Ω in bridge)
80×10-6 in strain p-p RTI(W/B,RANGE=1k,Without FINE,BV=2V,120Ω in bridge)
20×10-6 in strain p-p RTI(DC to 30kHz,RANGE=1k,Without FINE,BV=2V,120Ω in bridge)
Output OUTPUT1: ±10V,±5mA
OUTPUT2: ±10V,±10mA
Output adjustment OUTPUT2 ADJ(Can be independently varied continuously from 1 to 1/10)
Output monitor display 17-od LED display(OUTPUT1 monitor)
LED at both ends blink when voltage is greater than approximtely ±10.5 V.
Digitl display 4 1/2 digital display(OUTPUT2 monitor),Scaling display available with OUTPUT2 ADJ,
Displaying location of decimal point can be changed by using the bottom setting SW.
Key lock function Turning the key lock ON/OFF by pressing the key lock button approxmately for one second.
(Except CAL switch and BV selection switch)
Setting value saving Saving the value in flush memory(Can be held without back-up battery)
Resistance to vibration 29.4m/s2
Withstand Voltage
(Insulation resistance)
1 kVAC,1 minute,between each input terminal(A,B,C,D,E),output and housing case.
1.5 kVAC,1 minute, between AC power input and output or housing case(Includes serge resistor).
1 kVAC,1 minute,between DC power input and input.
500VAC,1 minute,between DC power input and output or housing case.
1.5 kVAC,1 minute, between AC power input and output or housing case(Includes serge resistor).
1 kVAC,1 minute,between DC power input and input.
500VAC,1 minute,between DC power input and output or housing case.
Ac power supply 85 to 132 VAC/180 to 264 VAC(Internal switch must be changed),9VA or less
85 to 132 VAC/180 to 264 VAC(Internal switch must be changed),10VA or less
DC power supply 10 to 30 VDC,6VA or less
10 to 30 VDC,7VA or less
Operating environmental conditions -10 to 50℃,Within 20 to 85%RH
without condensation
Outline dimension H143(±1.0)×W49.5(±0.5)×D253(±2.0)mm
*Excluding protrusion
Weight Within 1.4kg

Specifications for Model 5693 as DC Amplifier

Item Description
Input impedance Approx.10MΩ+Approx.10MΩ(DC)
Zero adjustment range ±10mV(RTI,BV=2V)
Balancing adjustment resolution Within ±1μV(RTI,RANGE=1k,without FINE,BV=2V)
Measuring range ±125mV(RANGE=50k,FINE=x2.5,BV=2V)
Gain 10,000x(RANGE=1k),5,000x(2k),2,000x(5k),1,000x(10k),500x(20k),200x(50k),Without FINE
Gain resolution ±0.1%
Common mode rejection ratio(CMRR) 70dB or more(50Hz or 60Hz) at 1kΩ balanced input
Maximum allowable input voltage ±8VDC or AC peak value
Common mode allowable input voltage ±5VDC or AC peak value
Internal calibrator Set value: ±0.01 to 99.99mV(±0.01 to 59.99mV at BV=2V)
Accuracy: ±0.2%rdg+5μV
Linearity Within ±0.01%FS
Stability Zero point: Within ±1μV/℃,±5μV/24 hours
Sensitivity: Within ±0.01%/℃,±0.05%/24 hours
Noise 80μVp-p(RTI,W/B,RANGE=1k(10,000x),without FINE,BV=2V)
20μVp-p(RTI,DC to 30kHz,RANGE=1k,without FINE,BV=2V)

Dimensional Out Line Drawing

Dynamic Strain Amplifier

Instruction Manual・・・・5683/5684(pdf,1.8MB), 5693(pdf,1.9MB)

All the instruction nanuals are of Portable Documents Formats. For reading the instruction manual, it is necessary to be installed yourself in the Adobe Reader of Adobe Company. Adobe Reader is the registered trademark of Adobe Systems Co., Ltd.

get_adobe_reader

Home | About Us | Contact Us | Site Map... | tuochi Automatic Instrument Co.,Ltd  Tel: 021-51085998  Fax: 021-51069009  Mail: zhang@yanatoo.com